Technical Specifications - PREMIER ART (Fully Automatic High Volume Fibre Tester)
Application Results
Length Characteristics
Upper Half Mean Length, Mean Length, 2.5 % Span Length, 50% Span Length, (all in inches and mm)
Uniformity Ratio, Uniformity Index (both in %)
Short Fibre index
Strength Characteristics Bundle Strength (in g/tex), Elongation (in %)
Fineness Characteristics Micronaire Value
Colour Characteristics
% Reflectance (Rd)
Degree of yellowness (+ b)
Colour grade
(based on Upland Old, Upland New, Pima and User defined grade charts)
Trash Characteristics No. of Trash particles
% Trash area
Leaf Grade (1. USDA  2. Based on user definition)
Maturity Characteristics Maturity estimate
Graphical Fibrogram
Force – Elongation curve
Application Range
Cotton Fibre samples subject to the following valid ranges
Upper Half Mean Length 16 mm to 40mm
Strength Fibres with UHML 18mm to 40 mm up to strength value 339.4 g f
Micronaire 2.5 to 6.0 (sample size : 8.5 to 11.5 grams)
Color Rd: 50 to 90; +b: 4 to 15
Trash Min particle size: 0.2 sq mm
Measurement Principle 
Length & length uniformity
Length & length uniformity Optical
Bundle strength & Elongation Constant Rate of Elongation
Micronaire Air flow
Color & Trash Optical
Configuration Options
Modules for testing length (L), strength (S), Mic (M), Colour (C) and Trash (T) in the following configurations:
a. 1 LS + M + C
b. 1 LS + M + C/T
c. 2 LS + M + C/T
d. 2 LS + M + C
Module Options   
Module Trash
Software BALESMART – Software for Bale Management
Hardware Barcode Reader
Testing Speed   
 Up to 150 samples per hour with one operator
One sample refers 1 LSM, 2 C/T for 1 LS Module and 2 LSCT, 1 M for 2 LS module
Calibration
Instrument calibrated using internationally accepted USDA Cottons.
At field calibration possible with any user defined cottons with known values
  
 
User Interface   
Reports All measured characteristics in numerical and graphical format
Front End Language English
Dimensions
Top View Front View
Power
Consumption 1.2 KW
UPS (on-line) 2 KVA, Single Phase
Non UPS Supply
(3 KW) min
3 phase
Compressed Air
Compressor + Drier 6 - 7 bar
Consumption 80 – 100 litres per min.
Ambient Conditions
Relative Humidity 65 ± 2%
Temperature 27 ± 2° C for tropical countries
20 ± 2° C for temperate countries
Note: Technical Specifications subject to change without notice